Chemical Indicators
PROCHEM SSW – Steam Class 5 Integrating Indicator
- Moving front (wicking) Integrator, no interpretation necessary
- Effective across the normal sterilization range
- Parallels the performance of a biological indicator
- Conforms to applicable portions of ISO 11140-1
- Catalog # 4140-CI-SSW (250 indicators/box)
- 5 year expiry from date of manufacture
PROCHEM SSI – Steam Class 4 Multivariable Indicator
- Color change from white to black within 3 minutes or longer at 132 - 135°C
- Non-reversible color change
- Measures time, temperature and presence of steam
- Instant visual confirmation saturated steam has penetrated pack or pouch
- Indicator perforated for use as one 8” strip or two 4” strips
- Conforms to applicable portions of ISO 11140-1
- Catalog # 4140-CI-SINT (250 indicators/box)
- 3 year expiry from date of manufacture
PROCHEM Dry Heat Indicator Labels – Class 4 Multivariable Indicator
- Color change from green to dark brown/black within 10 minutes at 160°C
- Non-reversible color change
- Adhesive backing allowing them to easily be applied to items
- Distinguishes between processed and unprocessed items
- Conforms to applicable portions of ISO 11140-1
- Catalog # 4140-CI-DHI (1,000 indicators/roll)
- 3 year expiry from date of manufacture
PROCHEM BD – Class 2 Bowie Dick Type Tests
- Preassembled, disposable test pack
- Distinct, non-reversible pink to black color change
- 3.5 minutes at 132°C in a pre-vacuum steam sterilizer
- Lead Free Ink on Indicator Sheet
- Conforms to applicable portions of ANSI/AMMI/ISO 11140-5
- Catalog # 4140-CI-BDTLF-20 (Box of 20 test packs)
- Catalog # 4140-CI-BDTLF (Case of 80 test packs)
- 3 year expiry from date of manufacture
PROCHEM BD121 – Class 2 Bowie Dick Type Tests
- Preassembled, disposable test pack
- Distinct, non-reversible pink to black color change
- 11 minutes at 121°C in a pre-vacuum steam sterilizer
- Lead Free Ink on Indicator Sheet
- Conforms to applicable portions of ISO 11140-1 and ISO 11140-4
- Catalog # 4140-CI-BDTLF-121-20 (Box of 20 test packs)
- Catalog # 4140-CI-BDTLF-121 (Case of 80 test packs)
- 3 year expiry from date of manufacture
Product Items List
Item Code | Product | Process | Quantity Per Unit | |
---|---|---|---|---|
4140-CI-SSW | PROCHEM SSW – Steam Type 5 Integrating Indicator |
Steam: 121ºC to 135ºC |
250 indicators / box | |
4140-CI-SINT | PROCHEM SSI – Steam Type 4 Multivariable Indicator |
≥3 min at 132°C to 135°C |
250 indicators / box | |
4140-CI-DHI | PROCHEM Dry Heat Indicator Labels – Type 4 Multivariable Indicator |
Dry Heat: ≤10 min at 160ºC |
1000 indicators / roll | |
4140-CI-BDTLF-20 |
PROCHEM BD – Type 2 Bowie Dick Type Tests
|
Pre-vacuum steam: 3.5 min at 132ºC |
20 test packs / box | |
4140-CI-BDTLF | PROCHEM BD – Type 2 Bowie Dick Type Tests |
Pre-vacuum steam: 3.5 min at 132ºC |
80 test packs / case | |
4140-CI-BDTLF-121-20 | PROCHEM BD121 – Type 2 Bowie Dick Type Tests |
Pre-vacuum steam: 11 min at 121ºC |
20 test packs / box | |
4140-CI-BDTLF-121 | PROCHEM BD121 – Type 2 Bowie Dick Type Tests |
Pre-vacuum steam: 11 min at 121ºC |
80 test packs / case |